Syringe Pump, AV-500X
AV-500X Syringe Pump
Syringe Size - 10, 20, 30, 50/60 ml
Applicable Syringe - Compatible with syringe of any standard
VTBI - 0.1-9999 ml 1000 ml in 0.1 ml increments, 1000 ml in 1 ml increments
Flow Rate
- Syringe 10 ml: 0.1-400 ml/h
- Syringe 20 ml: 0.1-600 ml/h
- Syringe 30 ml: 0.1-900 ml/h
- Syringe 50/60 ml: 0.1-1300 ml/h
- 100 ml/h in 0.1 ml/h increments
- 100 ml/h in 1 ml/h increments
- Syringe 10 ml: 400 ml/h
- Syringe 20 ml: 600 ml/h
- Syringe 30 ml: 900 ml/h
- Syringe 50/60 ml: 1300 ml/h
Accuracy - 2% (mechanical accuracy1%)
Infusion Mode
- Flow rate: ml/min, ml/h, Time-based
- Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h, etc.
KVO Rate - 0.1-1 ml/h (in 0.1 ml/h increments)
Alarms - Occlusion, near empty, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm, pressure sensor error, syringe installation error, syringe drop off
Additional Features - Real-time infused volume, automatic power switching, automatic syringe identification, mute button, purge, bolus, anti-bolus, system memory
Drug Library - Available
Occlusion Sensitivity - High, medium, low
Power Supply, AC - 110/230 V (optional), 50/60 Hz, 20 VA
Battery - 9.6 1.6 V, rechargeable
Battery Life - 7 hours at 5 ml/h
Working Temperature - 5-40℃
Relative Humidity - 20-90%
Atmospheric Pressure - 860-1060 hpa
Size - 314*167*140 mm
Weight - 2.5 kg
Safety Classification - Class Ⅱ, type BF